Adjustment

The following equipment is required to carry out the adjustment:
- the test set-up as paragraph 4.0 and
- the level pin.
Starting the Adjustment:
Press the service key through the hole at the back of the detector
(Item 7) with the level pin and switch-on the detector.
The display will be:
LCD-indication
After 8 seconds warming-up time the indication switches between
measuring value and nominal value in intervals of one second. Now
the detector should be warmed-up for about 10 minutes.
The kind of measuring value is marked by the number of segments
of the bar chart:
1 segment:
2 segments: Nominal value of zero (0)
3 segments: Nominal value of sensitivity (2.40 vol%)
4 segments: Nominal value limit of measuring range 2.
For a successful ajustment all steps (zero
adjustment, sensitivity adjustment) must be
carried out.
52
The service key must not be pressed
any longer. When being in the
adjustment mode the alarm levels and
the operation time will not be displayed.
The alarm and the operation signal is
blocked.
True value, i. e. the actual measuring value