7. Getting started
7.4
Camera settings -Hitachi HV D20
Various settings of the Hitachi HV D20 AM have
to be checked. To do this, a live image must fi rst
be generated on the screen. This is done with
the "Intellicam" program that is installed together
with the Matrox Meteor 2 frame-grabber card.
7.4.1 Displaying a live image
• Start the Intellicam program via the start menu
Start / Programs / Matrox Imaging / Intellicam.
As the Intellicam program references inter-
faces that are used for microdissection soft-
ware as well, you will get an error message if
you have already installed the microdissection
software. Just ignore this message by clicking
"OK".
• Now specify the camera confi guration fi le you
are using.
To do this, select the "Select Camera Confi gu-
ration" option in the "File" menu and open the
HV-D20.dcf fi le in the directory C:\Program
Files\Leica LMD\DCF.
(There may be differences due to different in-
puts during the program installation.)
If using a different camera, of course, you have
to open the matching DCF fi le.
36
• Press the camera icon in the toolbar to activate
the "Continuous grab" function.
You will now see a live image in the image win-
dow of the screen display.
Fig. 31
Keys for setting the camera