Error Messages
For a certain type of failure, the controller will self-diagnose the
error and the messages described in the following table are
displayed.
Tab. 1
MESSAGE
CHECK
CONNECTION TO
PUMP
PUMP WAITING
INTERLOCK
FAULT: PUMP
OVERTEMP.
FAULT:
CONTROLLER
OVERTEMPERATURE
Turbo-V 1000 ICE Controller User Manual / 87-900-905-01 (M)
DESCRIPTION
Wrong connection between the
pump and the controller.
The interlock signal of P1
connector is activated by an
interruption of the link between
pin 3 and 8 of J1 connector, or
because the external interlock
signal is open.
The pump upper bearing
temperature exceeded 60 °C.
The pump lower bearing
temperature exceeded 100 °C.
The controller transformer
temperature exceeded 90 °C.
Instructions for Use
Error Messages
REPAIR ACTION
Check connection between
controller and pump.
Press the START push-
button twice to start the
pump.
Reset the short circuit
between pin 3 and pin 8 of
J1 connector, or close the
external interlock signal.
Wait until the temperature
decrease below threshold
value.
Press the START push-
button twice
to start the pump.
Wait until the temperature
decrease below threshold
value.
Press the START push-
button twice to start the
pump.