Instructions for Use
Error Messages
MESSAGE
FAULT: TOO HIGH
L0AD
FAULT: SHORT
CIRCUIT
OVERVOLTAGE
FLOW METER
ALARM
204/268
DESCRIPTION
In normal operation, the current
drawn by the pump is higher
than programmed (8 A).
In normal operation (after the
starting phase) the output
connection is shorted (output
current higher than 16 A).
A failure has occurred in the
controller power supply section,
or the controller has received a
spurious signal.
Alarm relating to the purge
flow. The flow value remained
at less than the set threshold
for 10 seconds or more.
Turbo-V 1000 ICE Controller User Manual / 87-900-905-01 (M)
REPAIR ACTION
Check that the pump rotor
is free to rotate.
Press the START push-
button twice to start the
pump.
Check connections and
shortages between pump
and controller. Press the
START push-button twice
to start the pump.
Press the START button
twice to start the pump
again. If the message is
redisplayed, contact Agilent
for maintenance.
Check the correct operation
of the gas supply circuit.