Herunterladen Inhalt Inhalt Diese Seite drucken

Error Messages - Agilent Technologies TwisTorr 84 FS AG X3508-64001 Bedienungshandbuch

Inhaltsverzeichnis

Werbung

Error Messages

The controller will self-diagnose some error causes and will display
the following messages:
Tab. 1
MESSAGE
CHECK
CONNECTION
TO PUMP
WAITING
INTERLOCK
PUMP
OVERTEMP.
CONTROLLER
OVERTEMP.
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (D)
DESCRIPTION
Wrong connection between the
pump and the controller.
OR
The pump temperature is below
0°C.
The interlock signal of P1
connector is activated by an
interruption of the link between pin
3 and 8 of J1 connector, or
because the external interlock
signal is open.
The upper bearing/pump
temperature exceeds 60°C.
The controller environment
temperature exceeds 70°C.
OR
The controller radiator temperature
is above 60°C.
Instructions for Use

Error Messages

REPAIR ACTION
Check connection between
controller and pump.
Press the START push-button
twice to start the pump.
Connect pin 3 to pin 8 of
P1 connector, or close the
external interlock signal.
Wait until the temperature
decreases below threshold
value.
Press the START push-button
twice to start the pump.
Wait until the temperature
decreases below threshold
value.
Press the START push-button
twice to start the pump.
205/300
16

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Twistorr 84 fs ag x3508-64002

Inhaltsverzeichnis