Herunterladen Inhalt Inhalt Diese Seite drucken

Agilent Technologies TwisTorr 84 FS AG X3508-64001 Bedienungshandbuch Seite 232

Inhaltsverzeichnis

Werbung

17
Technical Information
Letter Protocol Description
CRC corresponds to the sum (with inverted sign) of all the
preceding bytes.
e.g., the START command "A" in ASCII code = 41 inverted it will
be: FF + 1 - 41 = BF.
Tab. 7
CHARACTER
"A"
CRC
"B"
CRC
"C"
CRC
"D"
CRC
"E"
CRC
"F"
CRC
"G"
CRC
"H"
"I "
CRC
"J"
CRC
232/300
"G" = Parameters reading
"H" = Parameters writing
"I" = Request for operating status
"J" = Request for numerical reading
"K" = Request for counters reading.
ASCII
41h
-
42h
-
43h
-
44h
-
45h
-
46h
-
47h
-
48h
49h
-
4Ah
-
TwisTorr 84 FS AG Rack Controller User Manual / 87-901-037-01 (D)
CONVERTED ASCII FOR CRC CALCULATION
-
BFh
-
BEh
-
BDh
-
BCh
-
BBh
-
BAh
-
89h
-
-
B7h
-
B6h