4. Cut an opening in the panel or industrial enclosure per the panel cutout dimensions shown in
5. Place the terminal through the cutout from the front and secure with the clamping brackets and
By design, the rear cover plate can produce outward force on the (extended) retaining brackets,
increasing the overall stiffness of the enclosure structure and providing additional assurance that
the panel mount unit will remain securely in place.
2.3.2.
Harsh Enclosure
The harsh enclosure is made of stainless steel with a front panel angle of approximately 38
degrees. The harsh enclosure is designed to rest on a flat surface such as a table or desk top, or it
can be mounted to a vertical surface with the mounting brackets included with the terminal.
2.3.2.1.
Desktop Mounting
When the IND570 terminal will be placed on a flat surface, the four rubber feet included with the
terminal should be adhered to the bottom of the enclosure to prevent sliding. Locate the four rubber
2-6
METTLER TOLEDO IND570 Weighing Terminal Installation Manual
inches and [mm] in Figure 2-8.
NOTE: The cutout dimensions of IND570 match those of IND560.
Allen-head screws. The screws should be tightened to a torque of 5 inch pounds (0.55 N-m).
NOTE: Once the Allen-head screws have been tightened, and the unit secured into place, the
rear cover plate of the IND570 panel mount unit may be difficult to remove and replace for
servicing. If this occurs, slight loosening of the Allen-head screws should allow for removal and
replacement of the rear cover for service purposes.
Figure 2-7: Front Panel Gasket
Figure 2-8: Panel Cutout Dimensions