5.3 Transducer plug-in detection
Before the module can be in active measurement mode for a selected chan-
nel, the module must have recognized that a transducer has been plugged in.
The module faultlessly detects when a transducer is plugged in or removed, if
plug-in detection has been implemented through a TEDS module or a jumper.
The jumper can only be installed on a 1-wire lead.
TEDS module (option)
1-TEDS-BOARD
or
Jumper
Fig. 5.3: Plug-in detection with jumper or TEDS module
The following leads are checked during plug-in detection:
DQ401
Voltage source Uin
1-wire TEDS
A2888−1.1 en/de
DATA
ye
bk
DQ430
Bridge excitation/sense lead
0-wire/1-wire TEDS
23
5
TEDS
4
5
TEDS
4
1-wire
DQ809
Measured value
−