e. Press Next to calibrate zero point.
f. Press Adjust to accept calibration.
g. Press Done to return to measurement.
5.
When complete, reinstall the sensor into the flow
housing.
6.
Cycle the sample flow on and off to remove any
retained bubbles. Restore flow to 200 – 500 mL / min.
7.
When the measurement has returned to its normal
value and is stable, deactivate the HOLD function in
the instrument or allow it to time out. See the trans-
mitter manual for instructions.