GB
3.4 Reset of subtotals 1-9 (from serial no. 0213-00100 2000)
Operation
3.5 Directions of measuring accuracy
Due to its construction the flow meter counts the volume of the liquid flowing through being calibrated for use with
AdBlue. The highest measuring accuracy will be reached with a constant flow rate.
3.6 Calibration
To improve the accuracy, the flow meter can be adjusted to the respective applications by calibration.
The calibration factor is either changed directly (from serial no. 023-0000 2000) or the flow meter calculates the
accurate correction factor after a comparative measurement between the displayed and the actual filled volume.
3.6.1 Calibration through comparative measurement
Filling a volume
Operation
Set the flow meter to one of the subtotals -9.
Fill the liquid into an exact measuring vessel e.g. 20 litres.
Switching into calibration mode
At the same time press for approx. 5 seconds
8
At the same time press for approx. 5 seconds
Display
Display