Inspiratory resistance
,
***
*****
Expiratory resistance
,
***
*****
PEEP generated by resuscitator
in normal use with added supply
gas flow
,
***
*****
Patient connector
Expiration connector
(for PEEP valve attachment)
Manometer port connector
Bag refill valve connector
*
Forward and backward leakage
O₂ inlet connector
14
Mark IV Baby
Mark IV Adult
w. oxygen reservoir tube:
0.6 cmH₂O at 5 l/min
4.0 cmH₂O at 50 l/min
3,7 cmH₂O at 50 l/min
w. oxygen reservoir bag:
0.8 cmH₂O at 5 l/min
4.9 cmH₂O at 50 l/min
1.3 cmH₂O at 5 l/min
2,2 cmH₂O at 50 l/min
4.4 cmH₂O at 50 l/min
w. oxygen reservoir tube:
< 2 cmH₂O
at 5, 10, and 15 l/min
< 2 cmH₂O
w. oxygen reservoir bag:
at 5, 10, and 15 l/min
2.5 cmH₂O at 5 l/min
(V
225 ml and 600 ml,
T
3.7 cmH₂O at 10 l/min
4.5 cmH₂O at 15 l/min
(V
20 ml, f 60)
T
Outside 22 mm male (EN ISO 5356-1)
Inside 15 mm female (EN ISO 5356-1)
30  mm male (EN ISO 5356-1)
Ø 4.2 +/- 0.1 mm
-
Inside 32 mm female
Not measurable
According to EN ISO 13544-2
Operation temperature limits
Storage temperature limits
Recommended long term storage in closed packaging at room temperature,
away from sunlight.
Notes:
– 10 cmH₂O = 1.0 kPa
– V
: Ventilation volume, f: Frequency (breath per minute).
T
Tested according to EN ISO 10651-4.
*
Higher airway pressure can be obtained by overriding the pressure-limiting valve.
**
***
At general test conditions according to EN ISO 10651-4:2009.
****
Values are approximate.
Maximum values
*****
f 20)
5.3. MRI Safety Information
The Ambu Mark IV is MR Conditional, and therefore may be safely used in the MR
environment (not inside the MR bore) under the following conditions.
Static magnetic field of 7 Tesla and less, with
Maximum spatial field gradient of – 10,000 G/cm (100 T/m)
Maximum force product of – 450,000,000 G
Use inside the MR bore may influence MR image quality.
RF-induced heating and MR image artifacts have not been tested. Any metallic parts are
fully encapsulated and do not have any contact with the human body.
Mark IV Baby
-18 °C to +50 °C (-0.4 °F to +122 °F)
*
-40 °C to +60 °C (-40 °F to +140 °F)
*
/cm (450 T
/m)
2
2
Mark IV Adult