Inspiratory resistance
,
***
*****
Expiratory resistance
,
***
*****
PEEP generated by resuscitator
in normal use with added supply
gas flow
,
***
*****
Patient connector
Expiration connector
(for PEEP valve attachment)
Manometer port connector
Bag refill valve connector
*
Forward and backward leakage
O₂ inlet connector
14
Mark IV Baby
Mark IV Adult
w. oxygen reservoir tube:
0.6 cmH₂O at 5 l/min
4.0 cmH₂O at 50 l/min
3,7 cmH₂O at 50 l/min
w. oxygen reservoir bag:
0.8 cmH₂O at 5 l/min
4.9 cmH₂O at 50 l/min
1.3 cmH₂O at 5 l/min
2,2 cmH₂O at 50 l/min
4.4 cmH₂O at 50 l/min
w. oxygen reservoir tube:
< 2 cmH₂O
at 5, 10, and 15 l/min
< 2 cmH₂O
w. oxygen reservoir bag:
at 5, 10, and 15 l/min
2.5 cmH₂O at 5 l/min
(V
225 ml and 600 ml,
T
3.7 cmH₂O at 10 l/min
4.5 cmH₂O at 15 l/min
(V
20 ml, f 60)
T
Outside 22 mm male (EN ISO 5356-1)
Inside 15 mm female (EN ISO 5356-1)
30 mm male (EN ISO 5356-1)
Ø 4.2 +/- 0.1 mm
-
Inside 32 mm female
Not measurable
According to EN ISO 13544-2
Operation temperature limits
Storage temperature limits
Recommended long term storage in closed packaging at room temperature,
away from sunlight.
Notes:
– 10 cmH₂O = 1.0 kPa
– V
: Ventilation volume, f: Frequency (breath per minute).
T
Tested according to EN ISO 10651-4.
*
Higher airway pressure can be obtained by overriding the pressure-limiting valve.
**
***
At general test conditions according to EN ISO 10651-4:2009.
****
Values are approximate.
Maximum values
*****
f 20)
5.3. MRI Safety Information
The Ambu Mark IV is MR Conditional, and therefore may be safely used in the MR
environment (not inside the MR bore) under the following conditions.
•
Static magnetic field of 7 Tesla and less, with
•
Maximum spatial field gradient of – 10,000 G/cm (100 T/m)
•
Maximum force product of – 450,000,000 G
Use inside the MR bore may influence MR image quality.
RF-induced heating and MR image artifacts have not been tested. Any metallic parts are
fully encapsulated and do not have any contact with the human body.
Mark IV Baby
-18 °C to +50 °C (-0.4 °F to +122 °F)
*
-40 °C to +60 °C (-40 °F to +140 °F)
*
/cm (450 T
/m)
2
2
Mark IV Adult