Electromagnetic compatibility (EMC) envi-
ronment
This equipment is for use in home healthcare
facility environment.
This equipment is not likely to cause any harmful
interference in nearby equipment; however, it
can be impaired by electromagnetic (EM) inter-
ference emitted by other equipment, which can
lead to malfunctions. Do not use this equipment
where there is a risk that other electrical and/
or electronic equipment can interfere with this
equipment.
This equipment complies with the following
standards:
IEC 60601-1-2:2014 Medical electrical equip-
ment - Part 1-2: General requirements for basic
safety and essential performance - Collateral
Standard: Electromagnetic disturbances - Re-
quirements and tests (EN 60601-1-2:2015)
EN 12182:2012, Clause 7, Assistive products for
persons with disability – General requirements
and test methods
Testing for EM disturbances (EM emission and EM immunity)
Test modes during emission tests:
Movement mode (battery). Actuator running in/out
Charging mode, actuator not running.
EM emission standards
Conducted and radiated RF emission CISPR 11
Harmonic distortion
IEC 61000-3-2
Voltage fluctuations and flicker
IEC 61000-3-3
EN
EMC - WARNINGS
Use of this equipment adjacent to or stacked
with other equipment should be avoided
because it could result in improper operation. If
such use is necessary, this equipment and the
other equipment should be observed to verify
that they are operating normally.
Use of accessories, transducers and cables
other than those specified or provided by
R82 could result in increased EM emission or
decreased EM immunity of this equipment and
result in improper operation. Several power
plugs are supplied with this equipment. Please
use the correct power plug suitable for your
country. Please chose among the supplied
power plug adaptor units: English, Australian,
EU, US/Japanese plugs.
Portable radio frequency (RF) communica-
tion equipment (including peripherals such as
antenna cables and external antennas) should
be used no closer than 30 cm (12 inches) to
any part of the heron w/power, including cables
specified by R82. Otherwise, degradation of the
performance of this equipment could result.
Classification/levels
Group 1 (the product uses RF energy only for its internal
function)
Class B (the product is suitable for use in all establish-
ments, including domestic establishments and those di-
rectly connected to the public low-voltage power supply net-
work that supplies buildings used for domestic purposes)
Class A.
No testing performed, as the rated power for this equip-
ment is less than 75 W. No limits are specified for equip-
ment with a rated power of 75 W or less.
Complies