Herunterladen Inhalt Inhalt Diese Seite drucken

B. Braun Aesculap Spine S4 FW692R Gebrauchsanweisung Seite 13

Perkutan-instrumente
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 18
Manual pre-cleaning with a brush
Phase
Step
I
Disinfectant
Cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
Phase I
Fully immerse the product in the cleaning/disin-
fecting solution for at least 15 minutes. Ensure
that all accessible surfaces are moistened.
Clean the product with a suitable cleaning brush
until all discernable residues have been removed.
Brush through all surfaces not accessible to visual
inspection, e.g. in products with hidden crevices,
lumens or complex geometries, for at least 1 min or
until no more residues can be removed, using a
suitable cleaning brush (TA007747 or TE654202).
Maneuver non-rigid components, such as adjust-
able screws, hinges etc during cleaning.
After cleaning, thoroughly rinse through these
components (at least five times) with the cleaning
solution, using a disposable syringe (20 ml).
Do not use metal cleaning brushes or other abra-
sives that would damage the product surfaces and
could cause corrosion.
T
t
Conc.
[°C/°F]
[min]
[%]
RT
>15
2
(cold)
RT
1
-
(cold)
Water
Chemical
quality
D–W
BBraun Stabimed; aldehyde-free,
phenol-free and QUAT-free
D–W
-
Phase II
Rinse/flush the instrument thoroughly (all accessi-
ble surfaces) under running water.
Maneuver
non-rigid
as adjustable screws, hinges, etc. during rinsing.
components,
such
11

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis