Herunterladen Inhalt Inhalt Diese Seite drucken

B. Braun Aesculap Spine activ C FW645R Gebrauchsanweisung/Technische Beschreibung Seite 19

Handgriff, fräserführung, probeimplantate
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24
Manual Pre-Cleaning with ultrasound and brush
FW870/FW871R/FW784R
Note
3
Clean reamer guide
, chisel guide
with locking pipes
11
in closed position (
move them during cleaning.
Stage
Step
I
Ultrasonic cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
Stage I
Clean the product in an ultrasonic cleaning bath
(frequency 35 kHz) for at least 15 minutes. Ensure
that all accessible surfaces are immersed and
acoustic shadows are avoided.
Clean the product with a suitable cleaning brush
until all discernable residues have been removed.
Use a suitable cleaning brush (TA011944) to clean
all surfaces which are not visible, e.g. in
instruments with concealed crevices, lumens or
complex geometries, for at least 1 min or until no
further residues can be removed.
Mobilize non-rigid components, such as set screws,
joints, etc. during cleaning.
After cleaning, use a 20-ml single-use syringe to
rinse thoroughly, for at least 5 times, these difficult
to access parts of the product.
Do not use metal cleaning brushes or other
abrasives that would damage the product surfaces
and could cause corrosion.
10
2
and handle
) and do not
T
t
Conc.
[°C/°F]
[min]
[%]
RT
>15
2
(cold)
RT
1
-
(cold)
Stage II
Water
quality
D–W
BBraun Stabimed; aldehyde-free,
phenol-free and QAV-free
D–W
-
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Chemical
17

Werbung