Herunterladen Inhalt Inhalt Diese Seite drucken

B. Braun Aesculap Spine activ C FW645R Gebrauchsanweisung/Technische Beschreibung Seite 18

Handgriff, fräserführung, probeimplantate
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 24
Aesculap Spine
activ C handle, reamer guide, chisel guide, trial implants
Manual pre-cleaning with a brush
FW789R–FW800R
Stage
Step
I
Disinfectant
cleaning
II
Rinsing
D–W:
Drinking water
RT:
Room temperature
Stage I
Fully immerse the product in the cleaning/
disinfecting solution for at least 15 minutes.
Ensure that all accessible surfaces are moistened.
Clean the product with a suitable cleaning brush
until all discernable residues have been removed.
Use a suitable cleaning brush (TA011944) to clean
all surfaces which are not visible, e.g. in
instruments with concealed crevices, lumens or
complex geometries, for at least 1 min or until no
further residues can be removed.
Mobilize non-rigid components, such as set
screws, joints, etc. during cleaning.
After cleaning, thoroughly rinse through these
components (at least five times) with the cleaning
solution, using a disposable syringe (20 ml).
Do not use metal cleaning brushes or other
abrasives that would damage the product surfaces
and could cause corrosion.
Stage II
Rinse/flush the product thoroughly (all accessible
surfaces) under running water.
Mobilize non-rigid components, such as set
screws, joints, etc. during rinsing.
16
T
t
Conc.
[°C/°F]
[min]
[%]
RT
>15
2
(cold)
RT
1
-
(cold)
Water
Chemical
quality
D–W
BBraun Stabimed; aldehyde-free,
phenol-free and QAV-free