Herunterladen Inhalt Inhalt Diese Seite drucken

Func-Cycl": Automatic Datalogging With Adjustable Cycle-Time - WIKA CPH6200-Ex Betriebsanleitung

Explosionsgeschütztes hand-held druckmessgerät
Vorschau ausblenden Andere Handbücher für CPH6200-Ex:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

Hand-Held Pressure Indicator
CPH6200-Ex (Atex-Version)
GB
18
4.2 Func-CYCL": Automatic datalogging with adjustable
cycle-time
The Logger cycle-time is adjustable (see Configuration). For example,
setting "CYCL" = 60: A measurement is made every 60 seconds.
When the slow measurement "rAtE-Slo" is chosen, additionally a low
power function is available: "Lo.Po".
If "Lo.Po" is on, the device only will take a measurement at the point of
time of the recording. In between the recordings the measuring shut's
down. This decreases the power consumption enormously and there-
fore is recommended e.g. for long time recordings.
Max. number of measurements: 9999
Cycle-time:
A measurement contains:
Slow measuring rate (rAtE-SLo):
- Current value at that data point
- Min-Peak, Max-Peak since last data point
Fast measuring rate (rAtE-FASt, -P.dEt):
- Average since last data point
- Min-Peak, Max-Peak since last data point
Starting logging:
By pressing ´Store/Quit´ again the recording will be initiated.
After that the display shows 'St.XXXX' for a short time whenever a
measuring is recorded. XXXX is the number of the data point 1 ... 9999.
If the logger memory is full, the display will show:
Recording will be stopped automatically.
If the Low-Power-Logger-function is selected („Lo.Po = on") the device
switches itself off as soon as the memory becomes full.