2. To enable the data subset number to be changed, press F1. The data subset number will
now be flashing. To enable the data sub number to be changed, press F1 again. The data
set number will now be flashing. Press F1 again to change the data id number.
3. Press the down arrow key (
key (
) to increment the enabled number. For storing data, the number can be set to
any value, overwriting existing data is allowed. For recalling data, the number can only
be set to used values.
Note: If you press the up or down arrow key (
decrements by one. To accelerate the increment or decrement function, press and hold
the up or down arrow.
Storing a Measurement
To store a measurement:
1. Press
2. Press F1 and use the arrow keys (
3. Press F2 to save the data.
• If memory is full, FULL will appear on the primary display. Press F1 to choose another
data identity, press
• If the memory is not full, the data will be saved, the tester will automatically exit
Memory mode and the display will revert back to the previous test mode.
• If the data identity has been previously used, the display will show STO? Press F2
again to store the data, press F1 to choose another data identity, press
exit Memory mode.
Recalling a Measurement
To recall a measurement:
1. Press
2. Press F3 to enter the Recall mode.
3. Use F1 and the arrow keys (
fields will be dashes.
1. Press F3 to recall the data. The tester display will revert to the Test mode used for the
recalled test data, however, the
Memory mode.
2. Press F3 to toggle between the data id screen and the recalled data screen to check the
recalled data id or to select more data to recall.
3. Press
Clearing Memory
To clear all memory
1. Press
2. Press F4. The primary display will show Clr?
3. Press F4 again to clear all memory locations. The Tester returns to the measurement mode.
) to decrement the enabled number or press the up arrow
to enter Memory mode.
to exit Memory mode.
to enter Memory mode.
) to set the data identity. If no data has been saved, all
to exit Memory mode at any time.
to enter Memory mode.
) once, the number increments or
) to set the data identity
icon still appears, indicating the tester is still in
21
to