h) The location of the cabinet will directly influence the noise level. This level will rise
considerably in small spaces with very reflective surfaces because of the proximity of the
cabinets to the walls, and where fan speed has risen to compensate for air currents in the
room or the resistance of clogged filters.
i)
Correct cabinet performance should be tested before putting it into operation if any
modification has been made to the room or to the air intake system which may
considerably change the airflow characteristics in the room.
3.3.

PRECAUTIONS FOR FIRST CONNECTION

Before connecting it to the mains supply, proceed to make a general cleaning to remove
any dust particles that may be accumulated during transportation. Follow the instructions
on cleaning and disinfection.
Before using it for the first time, run the cabinet for around 6 hours to purge the filters.
New filters give off a characteristic smell which tends to disappear in a short period of time.
WARNING
The cabinet should work at the voltage shown on the name plate. It should be connected to a
suitable socket base (minimum 16A) with an earthed plug. The cable section recommended
for the base is 1.5 mm
WARNING
The plug connection must be located in such a place to prevent it from being unplugged
intentionally or accidentally during operation, e.g., next to a wall or closed to another
equipment.
UM-BiOptima-EN-1907
2
.
EARTH
WIRE
11 / 39