3. INSTALLING INSTRUCTIONS

For a correct start-up procedure of the MSC, observe the following step-by-step instructions.
3.1.

TRANSPORT

The cabinet is supplied in an individual package and in the suitable way to prevent it from
being damaged.
This manual and the individual Quality Control certificate of each cabinet are
included in the scope of supply.
3.2.

PLACING ON SITE

a) Special care should be taken when unpacking the front glass and the control display.
b) Use the same packaging base for the transport to its final location and it is highly
recommended doing it with a pallet jack.
WARNING
Handle the pallet jack carefully to avoid turn over risks due to the high gravity center.
Never open the frontal before the cabinet is fully fastened to its support stand.
WARNING
We advise against the use of lift trucks for placing the cabinet directly by its base.
WARNING
For any handling during assembly or dismantling of the cabinet, appropriate safety measures
should be taken, such as wearing safety gloves or shoes.
c) Cabinets should be placed away from entrances, corridors, ventilation supply points, etc.,
which might affect airflow in the work area.
d) Ensure the cabinet is set perfectly on the table or workbench in order to prevent
vibrations.
e) To prevent the work area from overheating, we recommend that the room temperature is
always within the comfort margins for working.
The optimal comfort margins are:
Temperature: 18-28ºC
-
Relative humidity: 50% ± 20%
-
If any of these conditions is not fulfilled, optimum cabinet performance cannot be
guaranteed.
f)
When necessary, air vents in the room should be modified or speed of air entry reduced
so that air is not aimed directly at the cabinet, as this may produce turbulence.
Side air vents are not appropriate for rooms where cabinets are installed.
g) Due to the ceiling temperature affects air distribution, sites should be selected with
enough space. We recommend 250 mm minimum free space between the exhaust and the
ceiling.
UM-BiOptima-EN-1907