Aesculap Power Systems
Acculan® 3Ti attachments

10. Troubleshooting list

10.1 General malfunctions

Malfunction
Loud noise
Tool not moving
Deficient cutting
performance
Tool cannot be
coupled
22
Cause
Defective gearing
Attachment not coupled
properly
Defective gearing
-
Blunt tool
Wrong tool
Debris
Deformation
Finding
Loud noise
Attachment can not be
extracted from the motor
unit
Motor drive not turning;
blocked
Motor running, but not
driving the saw
-
Tool overheats
Incompatible tool
Debris on tool or in
coupling
Deformed tools/coupling
components
Remedy
Have product repaired by the
manufacturer
Couple the attachment
correctly, see Coupling and
uncoupling of attachments on
the drill and reamer motor
unit
Have product repaired by the
manufacturer
Have product repaired by the
manufacturer
see TA011869
Use a new tool
Insert appropriate tool
Clean the tool/coupling
Have product repaired by the
manufacturer