5. Commissioning, operation
2-channel version:
Storable data sets: 8,000
With slow measurements:
Sensor 1: Measured value
■
Sensor 1: Min.-Peak, Max.-Peak since last save
■
Sensor 2: Measured value
■
Sensor 2: Min.-Peak, Max.-Peak since last save
■
Difference sensor 1 - sensor 2: Measured value
■
CALC off
■
CALC
■
change of pressure (only for sensor connection 1)")
With fast measurements:
CALC off
■
CALC
■
change of pressure (only for sensor connection 1)")
Sensor 1: Arithmetic mean
■
Sensor 1: Min.-Peak, Max.-Peak since last save
■
Sensor 2: Arithmetic mean
■
Sensor 2: Min.-Peak, Max.-Peak since last save
■
Difference sensor 1 - sensor 2: Arithmetic mean
■
CALC off
■
CALC
■
5.4.5 "Rate of change of pressure (only for sensor connection 1)")
5.5.2.1 Starting logger recording
1. Press the [STORE/QUIT] button for 2 seconds.
⇒
Lo66 rvn
2. Press [STORE/QUIT] button once more.
⇒
Automatic logger recording started.
WIKA operating instructions, model CPH6300
Downloaded from
www.Manualslib.com
rAtE SLo
: Difference sensor 1 - sensor - 2: Min.-Peak, Max.-Peak since last save
"active": Estimated calculation and dummy values (see chapter 5.4.5 "Rate of
rAtE FASt
: Min.-Peak and Max.-Peak since last save
"active": Estimated calculation and dummy values (see chapter 5.4.5 "Rate of
: Difference sensor 1 - sensor - 2: Min.-Peak, Max.-Peak since last save
"active": Estimated calculation and dummy mean values (see chapter
With the measurement mode
is selectable:
Lo
.
Po
recording, a measurement only occurs at the respective logger time. This
reduces the current consumption considerably and is therefore mainly
recommended for long-term measurements (e.g. leak testing).
is displayed
manuals search engine
and
rAtE P.dEt
rAtE SLo
, a current-saving function
. If this is
oN
, it operates that, while the logger is
EN
27