Herunterladen Inhalt Inhalt Diese Seite drucken

Ottobock 1E91 Runner Gebrauchsanweisung Seite 19

Vorschau ausblenden Andere Handbücher für 1E91 Runner:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
5.4 Alignment
5.4.1 Bench Alignment
INFORMATION
The figures for the bench alignment can be found at the beginning of this
document.
INFORMATION
► For determining the overall system height of all prosthetic components
for bench alignment, add 5-10 mm to the socket-floor measurement to
compensate for the spring effect of the prosthetic foot. Use the larger
value when the patient's weight is at the upper limit of the maximum
permissible patient weight.
TT bench alignment
Required materials: 662M4 goniometer, 743A80 50:50 gauge, alignment
apparatus (e.g. 743A200 PROS.A. Assembly)
Perform the assembly and alignment of the prosthetic components in the
alignment apparatus according to the following specifications:
   
Sagittal alignment prosthetic foot
a–p position of the reference point to the alignment reference
line: See illustrations at the start of the document
   
Frontal alignment of prosthetic foot
Lateral rotation: 0° (see Page 21)
   
Sagittal alignment pivot point knee joint
a–p position of pivot point to alignment reference line:
-15 mm
   
Connect the prosthetic foot and prosthetic socket using the chosen
adapters. Be sure to follow the adapter instructions for use.
   
Sagittal alignment prosthetic socket
In proximal area centred on alignment reference line
Socket flexion: Individual residual limb flexion + 5°
   
Observe the abduction or adduction position.
Bench alignment process
19

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

1e93 runner junior

Inhaltsverzeichnis