Directions For Use
The REF 880 and 890 Series Video Colposcope is intended for use in the electromagnetic environment specified below. The customer or user of
the REF 880 and 890 Series Video Colposcope should assure that it is used in such an environment.
Immunity Test
IEC 60601
Test Level
Electrostatic discharge
± 6 kV contact
(ESD)
± 8 kV air
IEC 61000-4-2
Electrical fast transient/
±2 kV for power supply
burst
lines
IEC 61000-4-4
±1 kV for input/output
lines
Surge
±1 kV differential mode
IEC 61000-4-5
±2 kV common mode
Voltage dips, short
>95% dip in 0.5 cycle
interruptions, and
voltage variations on
60% dip in 5 cycles
power supply input
lines.
30% dip for 25 cycles
IEC 61000-4-11
>95% dip in 5 seconds
Power frequency
3 A/m
(50/60Hz)
magnetic field
IEC 61000-4-8
Electromagnetic Immunity
Compliance Level
Electromagnetic Environment – Guidance
± 6 kV contact
Floors should be wood, concrete, or ceramic tile. If floors are
covered with synthetic material, the relative humidity should
± 8 kV air
be at least 30%.
±2 kV for power supply
Mains power quality should be that of a typical commercial or
lines
hospital environment.
±1 kV for input/output
lines
±1 kV differential mode
Mains power quality should be that of a typical commercial or
hospital environment.
±2 kV common mode
>95% dip in 0.5 cycle
Mains power quality should be that of a typical commercial or
hospital environment. If the user of the REF 880 and 890
60% dip in 5 cycles
Series Video Colposcope requires continued operation during
power mains interruptions, it is recommended that the REF
30% dip for 25 cycles
880 and 890 Series Video Colposcope be powered from an
uninterruptible power supply or battery.
>95% dip in 5 seconds
3 A/m
Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical commercial or
hospital environment.
Specifications