Herunterladen Inhalt Inhalt Diese Seite drucken

Temperature Measurements - Trotec easy therm Bedienungsanleitung

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
Clean and hygienic
- There is no risk of anyone hurting themselves on glass or
coming into contact with toxic mercury after breakages .
- Highest possible levels of safety when measuring the tempe-
ratures of small children and babies .
- The tip of the thermometer can be cleaned quickly and ea
sily using a cotton wool cloth that has previously been
moistened with alcohol . The thermometer is now hygieni-
cally clean and can be used to measure the temperature of
other family members .
Fever alarm
10 short alarm tones indicate that the child or adult whose tem-
perature is being measured probably has fever .

5. Temperature Measurements

The digital infrared thermometer measures the infrared energy
which is emitted from the ear drum and the surrounding tissue .
This energy is collected through the lens and converted into a
temperature value . Temperature measurements which are ta-
ken at the ear drum are the most precise of all . Temperature
measurements which are taken exclusively from the surroun-
ding tissue can result In lower temperature values . This in turn
can lead to a fever being wrongly diagnosed .
EN
Using The Thermometer Correctly In Order To Prevent Wrong
or Inaccurate Readings
-
P
meter emits a tone and the scale with the temperature units starts
to blink . Pull the ear in question gently towards the back of the head
and then upwards to straighten the ear canal . Place the sensor firmly
(but not too firmly!) in the ear canal and press the start key . The sen-
sor must remain in the ear canal until a further tone is emitted . This
tone indicates that the temperature measurement is completed .
- Your digital ear thermometer has been clinically tested . The measure-
ment results it provides are proven as long as it is used according to
the operating instructions .
Operating Instructions – EasyTherm Ear Thermometer
- Place a clean, new protective cap over the
measuring tip before carrying out any measu-
rements in the ear canal .
1 . Take out a protective cap and place it in
the holding fixture of the applicator included
in the scope of delivery .
2 . Press the measuring tip of the thermometer
through the holding fixture of the applicator
so that the protective cap fully encloses the
measuring tip of the ear thermometer .
3 . Remove the cap after use and replace with
a new cap before carrying out further tempera-
ture especially when carrying out mea-
surements on different persons .
ress the on/off key to power up the thermometer . The thermo-
B - 4

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis