To clean and disinfect the system and transducer
1 Turn off the system by pressing the Power button.
2 Unplug the power cord from the outlet.
3 Remove the disposable transducer sheath, if applicable.
4 Disconnect the transducer from the system. Temporarily place it where it will not cross-contaminate clean
equipment or surfaces while you clean the ultrasound console.
5 Clean the exterior surfaces of the ULTRASOUND SYSTEM to remove any debris or bodily fluids. Use the
following procedure:
a Use either a pre-moistened wipe or a soft cloth dampened with cleaner or disinfectant. Choose a cleaner
from the list of approved cleaners.
Approved cleaners/disinfectants for the ultrasound system
Cleaner/disinfectant
SaniCloth AF3 (gray top)
SaniCloth Plus (red top)
PI-Spray II
1
For maximum effectiveness, the component being cleaned must remain wet with disinfectant for a minimum period of time.
2
Qualified for use as an intermediate-level disinfectant for mycobacteria. For a more complete list, refer to the cleaners and
disinfection document available at
b Remove all gel, debris, and bodily fluids from the system.
c With a new wipe, clean the system, including the display, by wiping from clean areas to the soiled areas.
This method helps to avoid cross-contamination.
d Observe the minimum wet contact time and refer to manufacturer's instructions. Monitor the system
for wet appearance. Re-apply with a new wipe if no longer wet.
e Allow the ultrasound console to air dry in a clean, well-ventilated space.
SonoSite Edge II Ultrasound System
2
www.sonosite.com
Minimum wet contact time
3 minutes
3 minutes
10 minutes
1
7