SonoSite Edge II Ultrasound System
User Guide Errata
Measurement references .................................................................................................................................. 1
Cleaning and disinfecting ................................................................................................................................. 4
Before getting started ....................................................................................................................................................................... 4
Storing the transducer .................................................................................................................................... 13
Transporting the transducer .......................................................................................................................... 14
The following content was either not included or incorrect in the system user guide; the revisions will be made
in the next update.
Measurement references
The following measurement references have been corrected.
Measurement publications and terminology
Effective Regurgitant Orifice (ERO) in mm
Oh, J.K., J.B. Seward, and A.J. Tajik. The Echo Manual. 3rd Edition, Boston: Philadelphia: Lippincott Williams and
Wilkins, (2007), 73-76, 210.
ERO = MV Flow Rate/MR Vel * 100
Ejection Fraction (EF), percent
Oh, J. K., J. B. Seward, A. J. Tajik. The Echo Manual. 3rd Edition, Philadelphia: Lippincott Williams and Wilkins,
(2007), 115-116.
EF = ((LVEDV - LVESV)/LVEDV) * 100%
where:
EF = Ejection Fraction
LVEDV = Left Ventricular End Diastolic Volume
LVESV = Left Ventricular End Systolic Volume
SonoSite Edge II Ultrasound System
2
1