4.2.2.98. Setup.Initialise
Setting of default values for the following areas:
ActMeth:
Config:
Silo:
Assembly: All values under &Assembly.
Setup:
All:
The action must be triggered with &Setup.Initalise $G.
4.2.2.99. Setup.RamInit
Initializes instrument, see page 110. All parameters are set to their default
value and error messages are cleared. The user and silo memories will be de-
leted.
Command only possible in the inactive state of the Coulometer.
4.2.2.100. Setup.InstrNo
Instrument identification for report output.
Set the value with &Setup.InstrNo $G .
4.2.2.101. Diagnose.Report
Output of the report containing the adjustment parameters. The Coulometer
has to be in its inactive basic state.
4.2.2.102. Diagnose.Simulation.Keycode
0
1
2
3
4
28
21
26
756/831 KF Coulometer, Instructions of Use
Setup.Initialise.Select
Current method. Parameters, calculations, and assignments for
the data output, operands C01...C19.
All values under &Config.
The silo memory is deleted. Same function as delete entire silo.
All values under &Setup.
Values of the entire tree (except silo and method memory).
Setup.InstrNo.Value
Entering a keycode is like pressing the corre-
8
16
sponding key. The keys have the following key-
codes:
9
17
10
18
11
19
12
20
5
25
13
27
29
24
4.2 Remote control commands
ActMeth, Config, Silo,
Assembly, Setup, All
serial number, 8 ASCII characters
23
15
6
$G
$G
$G
$G
0...29
22
95