4 OPERATION
4.4.4 pH Simulation
A pH (or T°) value can be simulated in this menu, allowing the user to test his system
without any liquid. The simulated value influences the current and pulse outputs and the
alarm relay.
ENTER
SIMUL
1)
SIMUL
1)
SIMUL
1)
Quit the sub-menu SIMUL by pressing
until the user enters another sub-menu.
4.4.5 Display of electrode voltage
Display of the instantaneous value of the electrode voltage.
ENTER
VOLTAGE
ENTER
END
PH SIMUL
Enter the pH value
T ° SIMUL
0......9
Enter the temperature value
0......9
-456 m V
E-35-
PH CONTROLLER 8205
ENTER
ENTER
ENTER
ENTER
or
. The simulation remains active
PH = 00.00
0......9
PH = 10.50
00.0 ° C
0......9
021 ° C