Disposal
The Papillon nasal mask, AeroFix II headgear,
and AeroFix plus tubing clip contain no toxic
substances. Dispose of the nasal mask in
accordance with applicable national laws and
regulations. Except where stipulated otherwise,
before being taken to a recycling center the
nasal mask should be cleaned in accordance
with the procedures described in this Instruction
Manual.
Warranty
A two-year warranty applies for manufacturing
defects. The warranty does not cover normal
wear and tear. For further details, see the MAP
Medizin-Technologie GmbH General Terms and
Conditions of Business.
MAP Medizin-Technologie GmbH shall not be
liable for any impairments to the safety,
reliability, or performance of the product caused
by:
• use other than that specified in the
Instruction Manual
• use in conjunction with other products not
designed for this purpose
• any technical alterations made to the
product.
The mask must always be cleaned and
disinfected before being returned for servicing.
On completion of the above procedure, dry
the mask and pack it in a securely fastened
transport bag.
Symbols
See Instruction Manual
Warning! Failure to follow this
instruction could pose a health risk.
REF
Part number
Lot number
Date of manufacture (month, year)
Washing temperature 30 °C
Do not use chlorine bleach
Do not iron
Do not spin-dry
Technical Data
Standards
CE labeling in accordance with EC
directive 93/42/EEC, class IIa
Quality Management: development,
manufacture, sale, and service certified in
accordance with ISO 13485:2003
quality management principles.
EN ISO 17510-2
(Sleep apnea respiratory therapy, Part 2: masks and
application accessories)
Validated Cleaning and Disinfection
The general suitability of the Papillon
effective cleaning and disinfection by the procedures
specified in this Instruction Manual has been established
by an independent accredited test laboratory.
Flow Resistance
Pressure drop measured with mask cushion size 2
at 50 l/min = max. 0.1 hPa
at 100 l/min = max. 0.4 hPa
Acoustic Pressure
30.8 dB(A), measured with artificial head at 10 hPa
according to EN ISO 17510-2
Tubing Connection
∅ 22 mm for elastomer tubing
Weight
Approx. 92 g
Therapeutic Pressure
4 to 20 hPa
Materials Used
Mask (latex-free):
Headgear (neoprene-free): nylon, lycra, spandex,
Cleaning of Headgear
Pressure/Flow Curve
Mean flow [l/min]
Pressure Units
1 hPa = 1 mbar = approx. 0.98 cm H
®
nasal mask for
silicone, PC
polyurethane, polyester
P [hPa]
O