4.6

Required Aids

Reprocessing step Items Needed

Precleaning

Soft plastic brush/soft cloth
4.7
Precleaning
Use water for precleaning, a non-fixing disinfectant if necessary.
1. Remove surface contaminants with a soft brush/cloth while
holding the product in a water bath and/or rinsing the prod-
uct under running water.
4.8

Cleaning and Disinfection by Machine

The washer/disinfector must have fundamentally certified ef-
fectiveness (e.g. CE marking according to DIN EN ISO 15883).
The detergent and disinfectant must be suitable for medical de-
vices made of plastic and metal and have a pH value between 5.5
and 11. Prohibited ingredients: organic solvents, oxidizing
agents, halogens, aromatic/halogenated hydrocarbons.
It is essential to follow the cleaning agent and disinfectant man-
ufacturers' instructions.
1. Place the product in a suitable rinsing basket carefully. In do-
ing so make sure the product is not touching other instru-
ments or parts of instruments.
2. Start a tested program with the following properties:
– Thermal disinfection (5 to 10 minutes at 194 to 199 °F
[90 to 93 °C] – with a tolerance in accordance with
DIN EN ISO 17665, A
– Final rinse with distilled or deionized water.
– Sufficient product drying.