OBJ_DOKU-175853-006.fm Seite 27 Donnerstag, 14. Juni 2018 11:06 11
Disinfectants must be rinsed off thoroughly after use.
Maximum air pressure during drying: 2 bar.
Maximum water pressure during rinsing: 2 bar.
During machine cleaning and disinfection, do not exceed 95 °C.
Observe the specifications from the manufacturer of the deter-
gent and disinfectant and in the chapter "Validated procedures
in overview".
During sterilization, do not exceed 138 °C.
Do not sterilize in hot air.
6.2
Reprocessing limitation
Frequent reprocessing can damage this product. Observe the
safety instructions for checking the product. With visible damage
or functional impairment, the product may no longer be used.
6.3
Required Aids
Reprocessing step Items Needed
Precleaning
Soft plastic brush/soft cloth
Manual cleaning/
Soft plastic brush/soft disposable cloth
disinfection
(low particle)
Compressed air gun (2 bar maximum)
Machine cleaning/
Protector (Art. No. 20195-194)
disinfection
Sterilization
Protector (Art. No. 20195-194)
27