Herunterladen Inhalt Inhalt Diese Seite drucken

Sirona CEREC Omnicam Gebrauchsanweisung Seite 28

Inhaltsverzeichnis

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 1
4 Disinfect the mirror sleeve of the CEREC Omnicam with the HLD set
Not approved in the USA
Approved in the USA
10
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
4
Disinfect the mirror sleeve of the CEREC
Omnicam with the HLD set
NOTICE
Fragment: mirror sleeve: no water in camera
Make sure that no water gets into the camera when rinsing the mirror
sleeve.
The complete process for high-level disinfection (HLD) is as follows:
Cleaning the camera
➢ Cleaning process after using the CEREC Omnicam Camera: Wipe
the camera with the appropriate mirror sleeve so that the superficial
impurities cannot harden and stick to the surface of the mirror sleeve.
Use a lint-free cloth as listed below which has first been soaked in
one of the cleaning products.
Agent approved for USA/not approved for USA
Alpro
Merz
Dürr
Kerr Corporation
Rinsing the exterior of the mirror sleeve
➢ Rinse the exterior of the mirror sleeve.
Sirona Dental Systems GmbH
● Minuten Spray classic
● Minuten Wipes
● Plasti Sept
● Plasti Sept Wipes
● Pursept-A
● FD 312
● CaviCide
● Cavi Wipes
D3492.201.02.02.02 01.2016
62 82 839 D3492

Werbung

Inhaltsverzeichnis
loading

Diese Anleitung auch für:

Cerec bluecam

Inhaltsverzeichnis