1 Caring for the CEREC Bluecam and prismatic tube
Removing the prismatic tube
Disinfecting Bluecam
4
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
Prismenrohr ab_Textspalte
A
1. Press the prismatic tube against the camera body.
2. Press detent A.
NOTICE
Risk of damaging the front lens or prism.
➢ Push the prismatic tube straight toward the front, do not tilt it.
3. Pull off the prismatic tube.
4. Replace protective cap (E).
Disinfecting
NOTICE
Do not spray the CEREC Bluecam with or immerse it in cleaning agents
or disinfectants!
➢ Disinfect the CEREC Bluecam (including the mirror sleeve) with a
cloth which has been soaked in the agent specified in the section
"Care, cleaning agents, and disinfectants" in the device Operating
Instructions or as described in the section "Disinfect the prismatic
tube of the CEREC Bluecam with the HLD set. [ → 6]" of these
Operating Instructions.
Sterilizing
A
Detent