3 Caring for the CEREC Omnicam and mirror sleeve
Removing the mirror sleeve
Disinfecting
8
Operating instructions for CEREC Bluecam and CEREC Omnicam Care and HLD (high-level disinfection)
3
Caring for the CEREC Omnicam and mirror
sleeve
Components of the Omnicam
CCT Cam
A
Press detent to release
B
Mirror sleeve
C
Sapphire glass (coated)
The CEREC Omnicam is a very sensitive optical device and must
Fragment camera care
therefore be handled with the utmost care. Protect the coated sapphire
glass and the camera window against scratches and clean them with a
lint-free cloth and ethanol (commercially available cleaning alcohol).
1. Press the mirror sleeve against the camera body.
Removing the Mirror Sleeve
2. Press detent A.
NOTICE
Risk of damaging the camera window or the coated sapphire glass.
➢ Push the mirror sleeve straight toward the front; do not tilt it.
3. Pull off the mirror sleeve.
Disinfecting
NOTICE
Do not spray the CEREC Omnicam or immerse it in cleaning agents or
disinfectants!
➢ Disinfect the CEREC Omnicam (including the mirror sleeve) with a
cloth which has been soaked in the agent specified in the section
"Care, cleaning agents, and disinfectants" in the device Operating
Instructions or as described in the section "Disinfect the mirror sleeve
of the CEREC Omnicam with the HLD set [ → 10]" of these Operating
Instructions.
Sirona Dental Systems GmbH
D
Camera window
E
Calibration set
62 82 839 D3492