CORIOLISMASTER FCX100, FCX400 SEALING OF DEVICES | IN/FCX100/FCX400/MID/OIML-XA REV. A
1

Purpose of the instructions

These instructions describe the process of sealing a device.
CoriolisMaster FCB400, FCH400
Sensor in integral mount design with dual-
compartment housing
Sensor in integral mount design with single-
compartment housing
Sensor in remote mount design with single-
compartment housing
Sensor in remote mount design with dual-
compartment housing
CoriolisMaster FCB400, FCH400, FCB100, FCH100
Sensor in remote mount design
CoriolisMaster FCB400, FCH400, FCB100, FCH100
Sensor in integral and remote mount design with
tower extension
A kit with a sufficient amount of seal wire and the required
parts is supplied for each of these versions.
1.1

Why seal?

If an electronic, electro-mechanical or mechanical device
needs to be used for legal metrology, any electronic,
electro-mechanical or mechanical components of this device
must be protected against tampering from outside.
Such protection is usually provided by sealing the device.
For this purpose, a sealing wire is inserted through the
provided sealing openings on the device components. The
ends of the wire are secured with a pinch seal.
The seal wire must be laid such that the appropriate device
components (cover, connection screws, etc.) cannot be
opened or removed without damaging the seal wire.
Undetected manipulation of the device is impossible as a
result.
2
Activating the hardware write
protection
WARNING
Page
Risk of injury due to live parts.
5
Improper work on the device can result in electric shock.
— Before opening the device, switch off the power
7
supply.
— Observe the applicable standards and regulations for
13
the electrical connection.
11
Microprocessors are used in all FCX100 and FCX400 series
devices.
The operating programs use parameters for respective
Page
requirements, data for specific nominal diameters, the zero
9
point and range of the sensor affected, for pulse valences
and much more.
Page
10
If the device has been configured for a specific application,
the parameters may no longer be changed. To prevent any
changes of the parameters during later operation, the write
protection switch is used.
Once configuration of the device has been fully completed,
the write protection switch must be set to the "Read Only"
position. The device can then no longer be configured on
the LCD indicator or through the communication interfaces.
By applying a seal, you avoid deactivation of the write
protection of the device during operation and prevent any
changes in the device configuration from being made.
EN – 3