4 Design and Functions
5 Transport, packaging and storage
5.1
Transport
GB
Check the CPH65I0 intrinsically safe hand-held pressure calibrator for any damage that may have
been caused by transport. Obvious damage must be reported immediately.
5.2
Packaging
Do not remove packaging until just before use.
Keep the packaging as it will provide optimum protection during transport (e.g. change in
installation site, sending for repair).
5.3
Storage
Permissible conditions at the place of storage:
Storage temperature: -20...+60 °C
■
Humidity: <85 % relative humidity (non-condensing)
■
Avoid exposure to the following factors:
Direct sunlight or proximity to hot objects
■
Mechanical vibration, mechanical shock (putting it down hard)
■
Soot, vapour, dust and corrosive gases
■
Store the CPH65I0 in its original packaging in a location that fulfills the conditions listed above. If
the original packaging is not available, pack and store the CPH65I0 intrinsically safe hand-held
pressure calibrator as described below:
1. Wrap the CPH65I0 in an antistatic plastic film.
2. Place the CPH65I0, along with shock-absorbent material, in the packaging.
3. If stored for a prolonged period of time (more than 30 days), place a bag, containing a
desiccant, inside the packaging.
16
WARNING!
Before storing the intrinsically safe hand-held pressure calibrator (following
operation), remove any residual media. This is of particular importance if the
medium is hazardous to health, e.g. caustic, toxic, carcinogenic, radioactive, etc.
WIKA Operating Instruction, Model CPH65I0