1.4 Service Life
WARNING!
Risk of Injury or Damage
The product has been tested for the service life
stated in this manual. Use of the product beyond
this time period may cause product damage and
injury.
– Only use the product for the service life stated
in this manual. DO NOT exceed the service life
of the product.
– Remove the product from service when the
service life has been met. DO NOT continue to
use the product.
– Perform all maintenance according to the
recommended schedule in this manual.
The mobile patient lift has an expected lifetime of 8
years when used in accordance with safety instructions,
maintenance intervals and correct use, stated in this
manual. The effective product lifetime can vary according to
frequency and intensity of use.
Actuator Service Life
Number of lifts per day
1–2
3
4
5
1078985-I
Service life of actuator
(in years)
10
9
6
5
Number of lifts per day
6
7–9
10–13
14–27
1.5 Warranty Information
Terms and conditions of the warranty are part of the general
terms and conditions particular to the individual countries in
which this product is sold.
Contact information for your local Invacare office is located
inside the back cover of this manual.
General
Service life of actuator
(in years)
4
3
2
1
7