Commissioning
3.5 Transducer Installation
10. Take either transducer and apply a continuous lengthwise 1/8-inch bead of coupling
11. Slide transducer into a Mounting Frames back end first aligning the angled edge of the
12. Tighten the transducer Clamping Screw to hold the transducer in place.
28
compound across the center of the transducer emitting surface.
Coupling Compound
Emitting Surface
Back End
Figure 3-6
Transducer
transducer with the angled edge of the Mounting Frame. Keep transducer from making
contact with the pipe until it butts up against the Mounting Frame stop.
installation procedure for the other transducer
Transducer
Angled Edge
F-Connector
.
SITRANS FUH1010 Quick Start Guide
Operating Instructions, 1/2009, CQO:QSG009 Rev A
Repeat