CAUTION
Orthosis not set upright during the charging process and for storage
In certain cases, this can lead to falling caused by:
Altered damping behaviour of the orthosis due to accumulated air in the hydraulic unit.
→ Store the orthosis in an upright position only.
→ Charge the orthosis in an upright position only.
→ Before using the orthosis, verify that the damping behaviour corresponds to the chosen
mode.
→ Take note of noises or inconsistent damping behaviour.
→ Walk 10 to 20 steps in order to activate the automatic bleeding process.

4.2.4 Notes on Applying the Orthosis

CAUTION
Using the orthosis without shoes
In certain cases, this can lead to falling caused by:
Slipping when walking on smooth floors (tiles), since the foot component of the orthosis
does not have a shoe profile on the sole.
Altered swing phase triggering behaviour due to incorrect heel height.
→ Only use the orthosis with suitable shoes and and with shoes that have the specified heel
height established at the initial fitting.
CAUTION
Tying tall footwear (above the ankle)
In certain cases, this can lead to falling caused by:
Transmission of incorrect forefoot load data to the electronics. By tying footwear too tight
above the ankle, the spring element is bent even while it is not under load, which can distort
important sensor signals for the control unit.
→ In case of footwear that extends above the ankle, ensure the laces are not tied to tight.
CAUTION
Foreign objects between the leg and orthosis
This can lead to pressure points on the leg caused by:
Foreign objects at contact points between the leg and the orthosis.
→ Smooth out wrinkles in the padding material.
→ Check the leg for pressure points.
CAUTION
Improper fit of the orthosis
In certain cases, this can lead to injuries due to falling or skin damage caused by:
Insufficient support/hold by the orthosis shells.
→ Remove and reapply the orthosis immediately.
→ Observe the instructions for putting on and removing the orthosis.
CAUTION
Pinching skin in the area of the closures
This can lead to injuries/congestion caused by: