Herunterladen Inhalt Inhalt Diese Seite drucken

Aim-TTI PL-P SERIES Bedienungsanleitung Seite 29

Programmable power supplies
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

On the serial interface the response is sent immediately if in non-addressable mode, or when
addressed to talk, if in addressable mode.
Example:
I<n>?
Sequential command
Operation complete message generated immediately after <rmt> is sent.
Returns the current limit for output <n> in Amps in <nr2> numeric format. The syntax of the
response is
On the serial interface the response is sent immediately if in non-addressable mode, or when
addressed to talk, if in addressable mode.
Example:
V<n>O?
Sequential command
Operation complete message generated immediately after <rmt> is sent.
Reads and returns the output voltage on output <n> in Volts in <nr2> numeric format. The syntax
of the response is
On the serial interface the response is sent immediately if in non-addressable mode, or when
addressed to talk, if in addressable mode.
Example:
This command is not available for output 3 and, if issued, will generate a command error.
I<n>O?
Sequential command
Operation complete message generated immediately after <rmt> is sent.
Reads and returns the output current form output <n> in Amps in <nr2> numeric format. The
syntax of the response is
On the serial interface the response is sent immediately if in non-addressable mode, or when
addressed to talk, if in addressable mode.
Example:
DELTAV<n> <nrf>
Sequential command.
Operation complete message generated immediately after execution.
Set delta voltage for output <n> to <nrf>. The value of <nrf> must be in Volts; no multipliers are
allowed. If the value of <nrf>, after rounding, is outside the range of the specified output an
execution error will be generated and the corresponding error number will be placed in the
Execution Error Register, 140-142 (Maximum delta voltage value exceeded).
28
If the set voltage on output 2 is 12.55 Volts the response to the command V2? will
be V2 12.55<rmt>.
If the current limit for output 2 is 1·000 Amps the response to the command I2?
will be I2 1.000<rmt>.
If the output voltage on output 2 is 12·55 Volts the response to the command
V2O? will be 12.55V<rmt>.
If the output current from output 2 is 0·934 Amps the response to the command
I2O? will be 0.934A<rmt>.
I<n> <nr2><rmt>
<nr2>V<rmt>
<nr2>A<rmt>

Werbung

Inhaltsverzeichnis
loading

Inhaltsverzeichnis