Herunterladen Inhalt Inhalt Diese Seite drucken

Aim-TTI PL-P SERIES Bedienungsanleitung Seite 19

Programmable power supplies
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

GPIB Parallel Poll
The power supplies offer complete parallel poll capabilities. The Parallel Poll Enable Register is
set to specify which bits in the Status Byte Register are to be used to form the
The Parallel Poll Enable Register is set by the ∗PRE <nrf> command and read by the ∗PRE?
command. The value in the Parallel Poll Enable Register is ANDed with the Status Byte Register;
if the result is zero then the value of
The instrument must also be configured so that the value of
during a parallel poll operation. The instrument is configured by the controller sending a Parallel
Poll Configure command (PPC) followed by a Parallel Poll Enable command (PPE). The bits in
the PPE command are shown below:
bit 7 =
bit 6 =
bit 5 =
bit 4 =
bit 3 =
bit 2 =
bit 1 =
bit 0 =
Example.
To return the RQS bit (bit 6 of the Status Byte Register) as a 1 when true and a 0 when
false in bit position 1 in response to a parallel poll operation send the following commands
The parallel poll response from the generator will then be 00H if RQS is 0 and 01H if RQS is 1.
During parallel poll response the DIO interface lines are resistively terminated (passive
termination). This allows multiple devices to share the same response bit position in either wired-
AND or wired-OR configuration, see IEEE 488.1 for more information.
Power on Settings
All remote control instrument settings are cleared at power up as shown in the following list. The
power on settings are also installed if the REMOTE/LOCAL switch is set to local.
Status Byte Register
Service Request Enable Register
Standard Event Status Register
Standard Event Status Enable Register
Limit Event Status Register (Then set to show new limit status)
Limit Event Status Enable Register
Execution Error Register
Query Error Register
Parallel Poll Enable Register
Registers marked thus are specific to the GPIB section of theinstrument and are of limited
use in an ARC environment.
18
X
don't care
1
1
Parallel poll enable
0
Sense - sense of the response bit; 0 = low, 1 = high
?
?
bit position of the response
?
∗PRE 64
<pmt>,
Voltage set to 0V for 30V supplies and 4V for logic supplies.
Current set to 1mA for 30V supplies and 1A for logic supplies.
Electronic output switches are off.
is 0 otherwise the value of
ist
then PPC followed by 69H (PPE)
local message
ist
is 1.
ist
can be returned to the controller
ist
= 0
= 0
= 128 (pon bit set)
= 0
= 0
= 0
= 0
= 0
= 0

Werbung

Inhaltsverzeichnis
loading