Alternating current (A/AC)
Range
600.0 µA
6000 µA
60.00mA
600.0mA
6.000A
10.00A
Specified measurement range: 5 - 100% of the measurement range
Overload protection 600 V; frequency range 45 - 400 Hz
Fuses: µA/mA = self-resetting PTC fuse 4x 160 mA, internal resistance approx. <10 Ω
10 A = High-performance FF12AH1000V ceramic fuse
≤ 6 A = continuous measurement, >6A = max. 10 s at intervals of 15 minutes
The multimeter may display 3 counts when a measurement input is open.
TrueRMS peak (Crest Factor (CF)) ≤3 CF over the entire range
TrueRMS peak for non-sinusoidal signals plus tolerance:
CF >1.0 - 2.0
+ 3%
CF >2.0 - 2.5
+ 5%
CF >2.5 - 3.0
+ 7%
Resistance
Range
600.0 Ω*
6.000 KΩ
60.00 KΩ
600.0 KΩ
6.000 MΩ
60.00 MΩ
600 V overload protection
Measurement voltage: Approx. 1.0 V, measurement current approx. 0.7 mA
*Accuracy for measurement range ≤600 Ω was calculated after deducting lead resistance from the REL function
Resolution
0.1 µA
1 µA
0.01mA
0.1mA
0.001A
0.01A
Resolution
0.1 Ω
0.001 KΩ
0.01 KΩ
0.1 KΩ
0.001 MΩ
0.01 MΩ
Accuracy
±(1.3% + 6)
±(1.6% + 4)
±(2.0% + 7)
Accuracy
±(1.3% + 3)
±(1.2% + 6)
±(1.6% + 4)
±(3.0% + 6)
73