Herunterladen Inhalt Inhalt Diese Seite drucken

Transport, Packaging And Storage; Commissioning, Operation - WIKA TSD-30 Betriebsanleitung

Temperaturschalter
Vorschau ausblenden Andere Handbücher für TSD-30:
Inhaltsverzeichnis

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

5. Transport, packaging and storage / 6. Commissioning, ...

5. Transport, packaging and storage

5.1 Transport
GB
Check the temperature switch for any damage that may have been caused by transport.
Obvious damage must be reported immediately.
5.2 Packaging
Do not remove packaging until just before mounting.
Keep the packaging as it will provide optimum protection during transport (e.g. change in installation site,
sending for repair).
5.3 Storage
Permissible conditions at the place of storage:
Storage temperature: -20 ... +80 °C
Humidity: 45 ... 75 % relative humidity
WARNING!
Before storing the instrument (following operation), remove any residual media. This is of
particular importance if the medium is hazardous to health, e.g. caustic, toxic, carcinogenic,
radioactive, etc.

6. Commissioning, operation

6.1 Mounting
Required tool: Open-ended spanner (spanner width 27)
The sealing faces at the temperature switch and the measuring point always have to be clean.
Only ever screw in, or unscrew, the instrument via the spanner flats. Never use the case as a working
surface.
14
WIKA operating instructions temperature switch model TSD-30

Werbung

Kapitel

Inhaltsverzeichnis
loading

Inhaltsverzeichnis