• Unpacking
- Unpack the device carefully
- Any damage should be notified immediately to the shipping
agent (post office, railway network or transport company)
• Scope of delivery
- RET control-visc heated magnetic stirrer
- Power supply cable
- Operating instructions
- Screwdriver (safety circuit)
- USB cable
• Use
- for mixing and/or heating liquids.
• Area of use (only indoors)
- laboratories
- pharmacies
The device is suitable for use in all areas apart from domestic areas
and areas directly connected to a low-voltage supply which also
serves domestic areas.
• R emove the threaded plug (P)
• R emove the protective cap from the
support rod
• P lace the washer between the hous-
ing and the nut
• S crew in the support rod manually
until it cannot be tightened further
• T ighten the nut using an SW17 open-
end spanner
• A ssemble the accessories using a
boss head clamp
Unpacking
Correct use
- schools
- universities
Assembling the support rod
- Protective cover: H104
- Magnetic rods: IKAFLON 30 and 40 mm
- Temperature sensor: PT 100.70
- O Ring
O-Ring
Please slide the O-Ring on the power cord plug to avoid ingress of
water into the connector.
The safety of the user cannot be guaranteed if the device is used
in conjunction with accessories which are not supplied or recom-
mended by the manufacturer, if the device is used contrary to the
intended purpose against the manufacturer's instructions or if the
device itself or PCB has been subjected to modifications by third
parties.
Note:
When using bath attachments with a diameter over 180 mm,
please use support rod in conjunc-
tion with an extension. (See "Ac-
cessories")
DANGER
Risk of tipping!
Make sure that the center of mass
of the attached device does not
protrude over the safe area indicat-
ed with a dotted rectangle.
safe area
22