RIA452
Installation instructions
Environment
Endress+Hauser
Current output:
Linearity
Resolution
Temperature drift
Output ripple
Voltage output
Linearity
Resolution
Temperature drift
10.0.5
Installation
Mounting location
Panel, cutout 92 x 92 mm (3.62x3.62 inch) (see 'Mechanical construction').
Orientation
Horizontal +/- 45° in every direction
Ambient temperature range
-20 to +60 °C (-4 to 140 °F))
Storage temperature
-30 to +70 °C (-22 to 158 °F)
Operating height
< 3000 m above MSL (9800 ft)
Climate class
As per IEC 60654-1, Class B2
Condensation
Front: permitted
Device casing: not permitted
Degree of protection
Front IP 65 / NEMA 4
Device casing IP 20
Shock and vibration resistance
2(+3/-0) Hz - 13.2 Hz: ±1.0 mm
13.2 Hz - 100 Hz: 0.7 g
Electromagnetic compatibility (EMC)
• Interference immunity:
To IEC 61326 industrial environments / NAMUR NE 21
• Interference emissions:
To IEC 61326 Class A
0.1% of full scale
13 bit
0.1%/10K (0.1%/18 °F)
10 mV at 500
for frequencies
0.1% of full scale
13 bit
0.1%/10K (0.1%/18 °F)
Technical data
50 kHz
107