MobileDiagnost 2.2
消息(英文)
The changed kV value
will lead to
clinically
questionable dose
compared to what was
defined as default.
If you really want to
proceed with this new
dose setting, select
YES.
If not, the kV value
will be reset to the
default value. In case
this was not intended,
check for any
contamination on the
touch screen and clean
it if necessary.
The changed mA value
will lead to
clinically
questionable dose
compared to what was
defined as default.
If you really want to
proceed with this new
dose setting, select
YES.
If not, the mA value
will be reset to the
default value. In case
this was not intended,
check for any
contamination on the
touch screen and clean
it if necessary.
22
消息(中文)
kV 参数更改为在临床上
存在问题的剂量值,而不
是最初定义的剂量值。
请检查您是真的需要继续
使用这个新的剂量设置,
还是在曝光前将 kV 参数
重置为所需的值。 如果
这不是您想要的操作,请
检查触摸屏上是否存在任
何污染,并在必要时进行
清洁。
mA 参数更改为在临床上
存在问题的剂量值,而不
是最初定义的剂量值。
请检查您是真的需要继续
使用这个新的剂量设置,
还是在曝光前将 mA 参数
重置为所需的值。 如果
这不是您想要的操作,请
检查触摸屏上是否存在任
何污染,并在必要时进行
清洁。
可能的原因
您将 kV 参数更改到了四
倍以上,或者是触摸屏上
存在污染。
您将 mA 参数更改到了四
倍以上,或者是触摸屏上
存在污染。
消息
可采取的措施
仅在有意更改时才可继续
操作,否则请恢复至更改
前。 如果这不是您想要
的操作,请检查触摸屏上
是否存在任何污染,并在
必要时进行清洁。
仅在有意更改时才可继续
操作,否则请恢复至更改
前。 如果这不是您想要
的操作,请检查触摸屏上
是否存在任何污染,并在
必要时进行清洁。