Herunterladen Diese Seite drucken

Hillrom Allen A-71100-SCH Bedienungsanleitung Seite 59

Werbung

Verfügbare Sprachen
  • DE

Verfügbare Sprachen

  • DEUTSCH, seite 422
INSTRUCTIONS FOR USE
3.7.1
Pendant
Removal
1.
Make sure the brakes are locked (the locked indicator on
the Allen Advance Table display is green).
WARNING:
Failure to remove power from the Allen Advance Table could cause patient fall,
injury, or equipment damage.
2.
Turn the Allen Advance Table off, and unplug it from the facility's power source.
3.
Unscrew the nut from the Allen Advance Table
connector.
4.
Remove the nut from pendant cord, and install the nut on
the new pendant cable. Make sure the ribbed end of the
nut is toward the connector end of the pendant cable
5.
Disconnect the pendant cable from the Allen Advance
Table connector.
Replacement
1.
Make sure the O-ring is installed in the Allen Advance
Table connector.
2.
Align the tab on the connector end of the new pendant cable with the tab on the
inside of the Allen Advance Table connector.
3.
Connect the pendant cable to the Allen Advance Table connector.
4.
Screw the nut on to the Allen Advance Table connector.
5.
Plug the Allen Advance Table in, and turn it on.
6.
Make sure all pendant functions operate correctly.
Document Number: 80028190
Version: G
Page 59
English
Issue Date: 03 FEB 2023
Ref Blank Template: 80025117 Ver. F

Werbung

loading