part nr 425083. Earth ground terminal.
1.2.c. Electrical and pneumatic
connections
The electrical power line connecting the machine
must be suitably sized in relation to:
• the electric power absorbed by the machine,
indicated on its data plate (Fig. 1);
• the distance between the machine and the elec-
tric hook-up point, so that voltage drops under
full load do not exceed 4% (10% when starting up)
of the rated voltage specifi ed on the data plate.
- The user must:
• fi t a power plug on the power supply lead in
compliance with the relevant safety standards;
• connect the machine to its own electrical con-
nection - A, Fig. 2 - and fi t a differential safety
circuit-breaker with 30 mA residual current;
• install protection fuses on the power line that
are suitably sized in accordance with the indica-
tions provided on the machine data plate (Fig. 1);
• connect the machine to an industrial socket;
the machine must not be connected to domestic
sockets.
An effective grounding connection is essential for correct operation of the machine.
Make sure the available pressure and the rendered capacity of the compressed air system are
compatible with those required for correct machine operation - see the "Technical Data" sec-
tion. For correct machine operation, the compressed air supply line must provide a pressure
range from no less than 8.5 bar to no more than 16 bar and guarantee an air fl ow rate greater
than the average consumption of the machine, which is equal to 160 Nl/min.
For the machine to function correctly, the compressed air delivered suitably must be
suitably treated (not above 5/4/4 in accordance with ISO 8573-1).