Herunterladen Diese Seite drucken

DeVilbiss Healthcare drive SAMSOFT 175 Bedienungsanleitung Seite 29

Werbung

Verfügbare Sprachen

Verfügbare Sprachen

USER MANUAL FOR
VIII-4) ROUTINE MAINTENANCE
(Monthly or as required; to be carried out by an authorised
person). Routine maintenance consists of checking the patient lift
functionality:
● With the patient lift on level ground, check the 4 castors are
touching the floor.
● When the brakes are engaged, it must not be possible to move
the patient lift (level and stable floor).
● Check the handles on the mast to ensure the boom and hoist
legs move smoothly and without noise.
● Handle the handles on the mast and check there is no play
between the boom and the mast and between the mast and the
hoist legs.
● Check the plastic covers and protective housings are not broken
and there are no cracks or missing parts.
● Check the condition of the connecting cables: power cable,
handset, and actuator. If a cable shows the slightest trace of
burning, wear, cut or if the wires are visible, quarantine the
patient lift immediately and have the cables replaced by the
technical service.
● The slings must be checked at least once a month, or more
regularly if used intensively, to ensure the seams are not undone,
the weft of the upholstery and slings is not visible and there are
no torn parts.
Never use a defective sling.
Only authorised service personnel or Drive DeVilbiss Sidhil
Service Technicians should carry out repairs or service
activities.
Failure to do so may result in the manufacturer's warranty
becoming void.
Worn or malfunctioning parts must be replaced with original
parts manufactured by Drive DeVilbiss Healthcare France.
Complete a load test before putting patient lift back into service.
VIII-5) PREVENTIVE MAINTENANCE
Annually or as required; to be carried out by an authorised person or Drive
Devilbiss Sidhil Service Technicians according with paragraphs 5. 1 , 5.2, and
5.3 below.
● It is a requirement to carry out preventative maintenance
annually.
● Battery pack to be changed approximately every 18 months,
depending on use.
● Annual maintenance must be carried out by technically qualified
personnel trained in the maintenance of electromechanical
equipment. Maintenance operations do not require any special
tools; adequate tools must be used according to the best
industry practices. Mechanical tightening, clearances, and
adjustments must be carried out in accordance with the best
industry practices.
VIII-5.1) MECHANICAL CONTROL
● Remove the cover from the hanger bar and observe the axle and
hanger bar assembly for wear:
SAMSOFT 175
Compliant assembly: the flange of the axle does not touch the
crossbar of the hanger bar.
Non compliant assembly: the flange of the axle touches the
crossbar of the hanger bar.
Caution: the axle and the hanger bar must be replaced
immediately. The patient lift must not be used until all
worn parts have been replaced and a verification has been
carried out before putting it back into service.
● If the assembly is compliant, disassemble the axle from the
hanger bar and clean any grease, debris or dust from the axle
and hanger bar.
● Inspect the axle, flange, and cross hole for signs of damage,
unacceptable wear or deformation.
A small amount of wear and tear may be seen on the inside of the
flange in contact with the lifting beam socket. This small amount
of wear is acceptable as long as the assembly is compliant (see
above); if in doubt, replace the axle immediately.
● Repeat the inspection on the hanger bar paying particular
attention to the condition of the socket and welds.
● Grease the axle and the socket of the hanger bar, taking care to
grease the flange part and the socket.
Note: Use a suitable metal-to-metal grease or contact Drive
DeVilbiss Healthcare France for any request.
● Carefully reassemble the assembly, ensuring everything is
secure and the hanger bar rotates correctly around the axle.
Caution: The axle and hanger bar must be replaced every 5
years or when the above mentioned wear occurs, whichever
comes first.
EN
Hanger bar crossbar
Hanger bar socket
Axle flange
29
29

Werbung

loading