USER MANUAL FOR
USER MANUAL FOR
III-2) UNBOXING
Make sure all the device elements have not been damaged
during transport. Make sure the accessories are present: the
lifting assembly (box with battery pack, actuator, handset), battery
charger, a user guide, and possibly a sling if it was ordered.
Note: Our device is mounted and tested in our manufacturing
site, but its packaging requires to be folded.
III-3) ASSEMBLY
CAUTION: before using the device, make sure all the
shafts, screws, and pins used during the installation are
perfectly engaged in their respective supports, and they
are correctly locked.
The SAMSOFT 175 is operational only if you use the sling
compatible with this device and if you respect the maximum
permissible load of 175 kg.
● Remove the structure from the box (possibly by cutting the
box corners), lay the structure on a flat floor, and then lock the
wheel brakes.
● In the lower part of the mast, remove the pin.
● Remove the safety shaft from the first hole of the mast.
22
SAMSOFT 175
SAMSOFT 175
Pin
2
nd
hole
1st hole
Safety shaft
● Straighten up the mast.
● Reposition the safety shaft in the second hole of the mast, and
then put the pin back.
● On the handset, press the lifting key to unfold the boom. The
ball joint linking the hanger bar to the mast automatically
disengages.
To avoid any risk of tearing, make sure the hanger bar
and particularly its hooks are clear of the cables and
mast in the vicinity during this unfolding step of the
boom.
● Tighten the adjusting screw (black wing screw).